ResMed AirCurve 11 ASV with HumidAir
Store Description
The ResMed AirCurve™ 11 ASV with HumidAir is a bilevel PAP device for people with central sleep apnea (CSA), obstructive sleep apnea (OSA), or a combination of sleep disordered breathing. This bilevel adjusts to one’s individual breathing patterns as they change during the night. Its algorithm, called adaptive servo-ventilation (ASV), detects when the user stops breathing and provides a breath. It treats therapy users with central sleep apnea, mixed apneas, or periodic breathing.
*ASV therapy is contraindicated in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea.
AirCurve 11 devices are for therapy users who benefit from greater pressure support. These bilevel machines include two pressure level settings: a higher pressure when one inhales, and a lower pressure that makes it easier to exhale.
Details
AirCurve 11 comfort and support features are designed to encourage/support adherence to therapy by creating a more natural and helpful bilevel therapy experience. Designed to help drive positive outcomes, AirCurve 11 empowers providers to act on therapy insights, stay connected with patients, and take bilevel care to the next level.
Features:
- AirView™
Using AirView, providers can stay connected to patients, view invaluable insights and reports, manage patient records and troubleshoot remotely. - myAir™
Through myAir, patients have access to Care Check-In and Personal Therapy Assistant to help them get started and stay on therapy. - AutoSet™
Adjusts in response to flow limitation, snoring and apneas, set between minimum EPAP and maximum IPAP limits to meet patients’ needs throughout the night. - HumidAir™
Integrated humidifier can be manually adjusted for temperature and humidity level - Easy-Breathe waveform
Intelligently recreates a patient’s breathing pattern so breathing feels more natural. - VSync
Monitors and compensates for unintentional mask leak, so patients receive their prescribed therapy pressure. - TiControl
Allows clinicians to set the minimum (TiMin) and maximum (TiMax) limits on the time the device spends in IPAP. - Trigger and cycle sensitivity
Triggers and cycles between IPAP and EPAP as changes in patient flow are sensed. - Touchscreen
Responsive and easy-to-use touchscreen makes navigating settings simple and intuitive. - Over-the-air updates
Deliver features directly to the device, giving patients and providers access to the latest in PAP technology.
Specifications:
- Intended delivered volume range: 200-2300 mL
- Maximum single fault steady state pressure: Device will shut down in the presence of a single fault if the steady state pressure exceeds 40 cm H2O (40 hPa) for more than 1 second
- Pressure measurement tolerance: ± 0.5 cm H2O (0.5 hPa) ±4% of measured reading
- Flow measurement tolerance: ± 6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
- Operating pressure range: EPAP: 4-15 cm H2O (4-15 hPa), Min PS: 0-6 cm H2O (0-6 hPa), Max PS: 5-20 cm H2O (5-20 hPa), Maximum pressure = 25 cm H2O (25 hPa)
- Dimensions with side cover: 3.72” H x 9.32” W x 5.45” D
- Device weight with side cover: 44 oz.
- Housing construction: Flame retardant engineering thermoplastic
- Hot plate material: Stainless steel
- Water capacity: 380 mL